The oxygen concentrator uses pressure swing adsorption technology. At normal temperature, the machine will continuously separate oxygen from the air.
The medical oxygen concentrator is composed of filter system, compressor, adsorption tower, electrical control system, humidifier system and air system.
It adopts the advanced transformation absorption principle. It separates the oxygen and nitrogen under common temperature and pressure, then get the medical oxygen which meets the medical standards.
Oxygen concentrator parts diagram
The Oxygen Concentrator is intended for individual use as an oxygen supplement device in a home or care facility. The patient is the intended operator.
The concentrator is an electronically-operated device that separates oxygen from the ambient air. It provides a high concentration of oxygen directly to you through a nasal cannula or other methods. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.
This equipment takes 120V ~ power source as a power source, air as raw material, adopts high quality and high-efficiency molecular sieve, and through the PSA method at room temperature, produces high purity oxygen conforming to medical standards, thus achieving sustainable and uninterrupted oxygen supply.
Oxygen concentrator manual
This user manual contains important information about the use of your concentrator and will serve as a reference for you as you use your concentrator.
Unscrew the cover of the humidifier. Add purified or distilled water into the humidifier bottle between the maximum and minimum water level lines. (If needed, add medicine into the water per the doctor’s recommendations.) Then screw the humidifier bottle onto the device.
Screw the humidifier bottle connector onto the cover of the humidifier. Then insert the humidifier into the holder on the front of the unit. Finally, connect the other end of the cannula to the oxygen outlet.
Plug in the power supply: Ensure that the power switch is off. Then plug the concentrator’s AC plug into a power outlet.
Turning the Concentrator On
Press the power switch. The display will read “HELLO”. At that time, the green, yellow, and red lights will turn on simultaneously indicating the machine is functioning as expected. After about 1 second, only the green light will remain on. After 4 seconds, the display will show the runtime.
To properly read the flow meter, locate the prescribed flow rate line on the flow meter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min line as prescribed.
Turning the Concentrator Off
Press the power switches and unplug the concentrator from the power outlet.
Oxygen Concentrator Working Principle
A portable oxygen concentrator receives air, purifiers it, and then distributes the air.
How oxygen concentrator works?
An oxygen concentrator uses the ambient air, which is normally 80% nitrogen and only 20% Oxygen.